A “cosine corrected”
sensor is designed to maintain its accuracy when
radiation comes from different angles. For pyranometers,
the test of cosine response is to measure extreme zenith
angles. The cosine response of Model SP-110 at 75˚ is ±
5%, see diagram at right and graph below. Zenith angles
greater than 75˚ contribute less than 3% of daily
radiation.The ideal cosine response in a vacuum is shown in red in the
graph below. The CM21 and Apogee SP-110 deviate from the
ideal due to humidity and particulate matter in the
atmosphere.

This graph
shows typical
Apogee cosine response relative to the K&Z CM21.
Apogee has devoted significant resources to improving the cosine
response of our pyranometers. These efforts have been
rewarded by independent verification from the Broadband
Outdoor Radiometer Calibration (BORCAL) test results at
the National Renewable Energy Laboratory (NREL) in
Boulder, Colorado.
In June of 2005 NREL confirmed that
all six replicate Apogee sensors (Model SP-110) exceeded
our specification for cosine response (±1% at 45° Zenith
angle, ±5% at 75°). Results of these tests are available
from NREL and are shown here
along with a sampling of
three thermopile (black body) pyranometers and two other
silicon cell pyranometers. These tests have also shown
that the CM21 has excellent cosine response (shown
here
). NREL normalizes all of the cosine error data to
zero error at 45 degrees, which is the approximate
average sun angle at most latitudes.
Because of its good cosine response, the
SP-110 reads
accurately throughout the day as the sun angle changes
from dawn to solar noon to dusk. This accuracy is
maintained as sun angles change from summer to winter.
Sensors with poor cosine response can be calibrated so
that they will accurately measure daily total radiation
when the sun angle is the same as the day of
calibration, however, they will not be accurate at
different times of the year.
This figure
shows the magnitude of seasonal changes in
sun angle in Logan Utah. Seasonal changes result in a
23.5o variation in solar angle between the solstice and
equinox and 47o between the winter and summer solstices.
For example, the zenith angle in Logan, Utah at solar
noon varies from 18.3o on June 22 to just 65.2o on
December 22. Remember that 0o is directly overhead.
Cosine response is
synonymous to the term Lambertian response. Lambert’s
Cosine law states that radiation intensity on a flat
surface decreases as the angle of the surface decreases
from perpendicular (normal or 0˚ zenith angle). This is
expressed as: Eθ = E * cos(θ)
All radiation
sensors have some azimuth error, which shows up as the
difference between the am and pm response. This error is
typically smaller than the cosine error. To minimize
azimuth error we calibrate Apogee sensors with the lead
wire pointing north (in the Northern hemisphere) and we
recommend mounting the sensor with the wire pointing
towards the nearest pole. Most of the azimuth error is
thus corrected in the calibration.
The castle design. A flat
sensor surface (without cosine correction) reflects
radiation at low angles and under-weights low angle
radiation. A sensor with a raised white diffusion disk
over-weights low angle radiation. The traditional
approach to achieving a good cosine response is to build
a sensor with a raised, white disk, and then add a
raised wall around the perimeter to block low angle
radiation (this is called the castle design). This is an
effective design, but it traps water and dust, which
block light and result in low readings. The Apogee
sensor uses a domed top to repel water and dust. This
makes the sensor self cleaning. Accurate cosine response
is achieved by having just the right amount of curvature
on the dome, as well as using an appropriately opaque
diffuser.